166
Data Management Software & Interface
Data Apex
III. Instruments
U-PAD is a miniature external measuring unit
designed for acquisition of data from any chro-
matograph for the CSW32 station.
U-PAD uses the USB communication channel
and thus is directly powered from the PC.
The unit can be connected or disconnected any-
time without the necessity of disconnecting the
computer.
The unit contains two independent channels with the latest
24-bit A/D converters capable of higher eective resolution es-
pecially at higher frequencies and at lower voltage ranges. The
input connector is compatible with the internal INT7 board.
Noise free resolution for various acquisition
speeds and input ranges:
Technical data:
Converter type 24-bit integrating, sigma - delta
Inputs dierential, isolated
Number of
channels
2
Input range ± 156, 1 250 or 12 560 mV
Input impedance 1 MOhm
Integration
frequency
6.25, 12.5, 25, 50 and 100 Hz (50 Hz rejection)
7.5, 15, 30, 60 and 120 Hz (60 Hz rejection)
Non-linearity < 0.0015 %
Number of digital
inputs/outputs
2/2 (TTL level), direct and isolated
Dimensions 100 x 55 x 16 mm
Weight 110 g
Power supply from the PC via the USB cable
Internal memory none
12 560 mV 1 250 mV 156 mV
6.25 Hz z bit 22 bit 20 bit
12.5 Hz 21 bit 21.5 bit 20 bit
50 Hz 20.5 bit 21 bit 19 bit
100 Hz 20 bit 20 bit 18.5 bit
U-PAD